Type: | Power Meter |
---|---|
Wiring Devices: | ODF |
Condition: | New |
Connector Type: | FC/PC (Interchangeable Sc, St) |
Application: | FTTH FTTB FTTX Network |
Product Name: | Hand-Held FTTH Power Meter |
Customization: |
---|
Suppliers with verified business licenses
Wavelength range | 800~1700 |
Connector | FC/SC/ST(ST apolegamy ) |
Type of Detector | InGaAs |
Power measurement range | -70~+3 or -50~+20 |
Uncertainty | ±5% |
Standard wavelength (nm) | 850,980,1300,1310,1490,1550,1625 |
Display resolution | Linear indication:0.1% Logarithmic indication:0.01dBm |
working temperature (ºC) | -10~+60 |
Storage temperature (ºC) | -25~+70 |
Auto off time (min) | 10 |
Battery Working Time (h) | ≥48 |
Instrument size (mm) | 190 *100 *48or165*75*28 |
Power Supply | No. 5 battery *3 or rechargeable lithium battery |
weight (g) | 300/400 |
Function | Push-button | Comments on a form |
Increase 0.05dB | dB | *1.01158 |
Reduce 0.05dB | LIGHT | *0.99855 |
Preservation | ZERO | Press the boot button(Store to flash) |
Switching wavelength | ||
Restore factory settings | +ZERO | Display "FAC", The default value is also read into the array. |
Malfunction Tips | Cause | Solution |
LCD monitor weak | Insufficient power | Charge or change battery |
Could not display when starting | Lack of power or other | Restart machine or charge, change battery |
abnormal data display of LCD | Adapter fault, dirty or be locked | Reconnect adapter and clean sensor |
Suppliers with verified business licenses